General probe specifications:
Parameter | Value |
---|---|
Resonator Frequency (empty) | 36 GHz |
Resoantor Frequency (with sample capillary) | 34 GHz |
Microwave Conversion Factor (critically coupled) | > 12 G/sqrt(W) |
Bandwidth (critcially coupled) | Q ~ 300 - 400 (85 - 113 MHz) |
Microwave Conversion Factor (overcoupled) | > 5 G/sqrt(W) |
Bandwidth (critcially coupled) | Q < 85 (> 400 MHz) |
Operating Temperature | 4 K to RT |
Sample Access
Currently, the Bridge12 QLP is only available for samples with an OD of up to 1.6 mm.
Resonator Model | Description |
---|---|
QLP-1.6 | Resonator height: 4 mm Maximum Sample OD: 1.6 mm |
QLP-2.5 * | Resonator height: 4 mm Maximum Sample OD: 2.5 mm |
* Resonator currently under development. Please contact info@bridge12.com if you are interested in this specific resonator model.