This is the multi-page printable view of this section. Click here to print.

Return to the regular view of this page.

Systems

Welcome to the online documentation of our systems for EPR, NMR and DNP spectroscopy. Bridge12 does not distribute paper manuals with the equipment to ensure our documentation is always up to date. If you have questions, or suggestions for edits please contact us at info@bridge12.com.

Important

Please read the documentation for you product carefully to avoid any damages to the Bridge12 system.

1 - Spectrometers

Systems for EPR, NMR, and DNP Spectroscopy

1.1 - X-Band IF

Bridge12 X-Band IF System for Pulse EPR Spectroscopy

The Bridge12 X-Band IF system is at the heart of all Bridge12 EPR spectrometers. The system is highly modular. For X-Band EPR spectroscopy only a high-power microwave amplifier, digitizer and AWG is required.

The system can be completely controlled using SpecMan4EPR.

1.1.1 - X-Band IF Backpanel Connections

Backpanel connections of the X-Band IF system.

Below, please find a list and description of the backpanel connectors of the X-Band IF system. Please refer to the figure below for the location of connections.

Warning

Whether a connector is used for an input or output signal is labeled in the table below. Please make sure the user makes the correct connections. Wrong connections can lead to permanent damages of the system.

RF Connections

Please refer to the table below for a description of the back panel connectors of the X-Band IF system. The function (I/O - input/output) of the connector is indicated in the second column.

Information

Make sure all connections to the back panel connectors are properly tighten. SMA connections should be tightened using a torque wrench (recommended torque 10 Nm).

Side Panel

Connector I/O Type Description
SIG I SMA Input signal for the receiver.
LO2 O SMA LO2 output of the microwave synthesizer (channel B)

Receiver (RCVR)

Connector I/O Type Description
RCVR I O SMA I channel of the quadrature receiver channel. This is typically referred to as the real signal of the detected EPR signal.
RCVR Q O SMA Q channel of the quadrature receiver channel. This is typically referred to as the imaginary signal of the detected EPR signal.
RCVR IF I SMA IF input of the RCVR. For X-Band operation, this should be connected to the LO1 output of the microwave synthesizer (channel A). Suggested Cable

Pulse Forming Unit (PFU) and AWG Input

Connector I/O Type Description
PFU TX O SMA Output of the PFU. This signal is typically routed to the high-power amplifier or the active multiplier chain (AMC) in high-field/high-frequency EPR spectrometers.
PFU BLNK I SMA Blanking gate of the PFU. TTL logic. Active high. This signal has to be connected to the pulse programmer.
AWG I I SMA I channel input of the PFU. This signal needs to be connected to the output channel of the AWG.
AWG Q I SMA Q Channel input of the PFU. This signal needs to be connected to the output channel of the AWG.
AWG AUX I SMA Auxillary input of the IQ mixer LO channel.

Synthesizer and Monitors

Connector I/O Type Description
AUX IN I SMA Auxillary microwave signal input. This signal is combined with the microwave signal generated by the IQ mixer of the PFU and allows the user to inject an additional, user-created microwave signal.
AUX OUT O SMA LO signal of the LO input for the IQ mixer of the PFU. This signal is similar to LO1.
LO1 O SMA LO1 signal of the X-Band IF system (channel A of the synthesizer).
LO2 O SMA LO2 signal of the X-Band IF system (channel B of the synthesizer).
RCVR O SMA Receiver input monitor.
PFU O SMA PFU output monitor.

Reference Clock and External IO

Connector I/O Type Description
10 MHZ O SMA Oven-controlled 10 MHz reference clock signal. Output power level 10 dBm.
VCA O SMA 0 - 5 V DC signal. The output level of this signal can be controlled through the software. This signal is typically used for the Voltage Controlled Attenuator (VCA) of a high-frequency AMC.
EXT O D-Sub 9 pin d-sub connector. Pin Out
IO O D-Sub 9 pin d-sub connector. Pin Out

Other Connections

Side Panel

Connector I/O Type Description
12 V O LEMO 12 V output. This supply voltage can be used to power up external equipment. Do not exceed 200 mA. Mating Connector

Other Connectors

Connector I/O Type Description
EXT PWR O M12 Power supply e.g. for a frequency extension. Available voltages: -12 V, -5 V, 5 V, 12 V, 15 V
PWR I Amphenol Power inlet for the X-Band IF system. Please only use the power supply supplied with the system to avoid permanent damages.
PLS I/O USB USB connection from the X-Band IF to the pulse programmer. This USB port is connected to the internal USB hub of the X-Band IF system.
PC I/O USB USB connection to remote PC.
GND n/a STUD A ground (GND) post is located in the lower left corner of the back panel.

Warning

Only use the power adapter that came with the X-Band IF system to power up the instrument.

Failure to use the correct power adapter can lead to permanent damage of the system.

If you are unsure about the power adapter, please contact Bridge12 at support@bridge12.com

1.1.2 - Digital Demodulation

How to use digital demodulation with the X-Band IF system.

The Bridge12 X-Band IF system supports digital demodulation. Digital demodulation allows for easy removal of baseline artifacts and will result in a much cleaner signal detection. Instead of down-converting the EPR signal to DC level, the LO frequency is slightly offset and the signal is detected at a frequency of e.g. 200 MHz. The exact frequency depends on the sampling rate and input bandwidth of the digitizer (or oscilloscope).

Information

Digital demodulation is highly recommended for the X-Band IF system. Even an offset of just 20 MHz will greatly remove many artifacts from the baseline.

The example below shows a digitally demodulated signal of a 2-pulse Hahn echo of a sample of BDPA in polystyrene, recorded at Q-band frequencies.

We recommend using an intermediate frequency of about half the bandwidth of the digitizer. For example, if the digitizer has an input bandwidth of 400 MHz, we recommend to choose an intermediate frequency of 200 MHz.

To acquire the EPR signal at a This offset can be achieved in two ways:

1. Offsetting the AWG Frequency

The simplest way to use digital demodulation is by offsetting the frequency for the AWG generated microwave pulses. Instead of generating a rectangular pulse at DC level, the user can create a pulse at a frequency of e.g. 200 MHz. To make sure the frequency of the microwave pulse is within the bandwidth of the resonator, the microwave frequency needs to be lowered by the same amount. For example, to generate a microwave pulse at 9.8 GHz with an offset of 200 MHz:

  1. Set the LO1 frequency of the synthesizer to 9.6 GHz (9.8 GHz - 0.2 GHz)
  2. Set the AWG frequency to 200 MHz
  3. Set the demodulation frequency in SpecMan4EPR to -200 MHz.

2. Offsetting the Receiver Frequency

If pulses are not created by an AWG but for example by a pulse generator, digital demodulation can still be used. However, in this case, both microwave synthesizers, LO1 and LO2, have to be utilized.

To use digital demodulation with DC pulses follow these steps:

  1. Connect the RCVR IF to the LO2 output.
  2. Set the LO1 frequency on-resonant with the resonator frequency, e.g. 9.6 GHz.
  3. Set the LO2 frequency to 9.4 GHz. That way the echo signal has to be demodulated using a frequency of 200 MHz (9.6 GHz - 9.4 GHz = 0.2 GHz).

Demodulating the Signal (Demodulating the Signal)

Once the signal is digitized it has to be digitally demodulated.

In SpecMan4EPR

Demodulating the EPR signal in SpecMan4EPR is very straight forward and simple. Simply make sure the Demodulation Frequency is set to the correct value and the spectrum is automatically demodulated when the data is acquired.

Manually

If the X-Band IF is used manually, demodulation must be performed by the user during post-processing of the EPR data. This can be conveniently done by using the demodulate function in DNPLab.

DC Detection

To down-convert the EPR signal to DC level, connect the LO1 output of the synthesizer to the IF input of the receiver and make sure the frequency offset for the AWG pulses is set to 0 MHz.

1.1.3 - Example Configurations

Example configurations for different frequency bands.

The Bridge12 X-Band IF system can be used to operate EPR spectrometers from S-Band (2 GHz) to the millimeter regime (> 400 GHz).

Below, please find example configuration for different operating frequencies.

1.1.3.1 - Example: Pulsed X-Band Operation

Example configuration for pulsed X-Band EPR spectroscopy.

A typical configuration of the X-Band IF system for pulsed X-Band spectroscopy is shown in the figure below. Connections that are not required are greyed out.

Connector Description
RCVR IF
to LO1
Connect a microwave cable between the LO1 synthesizer output and the RCVR IF input. This has to be a microwave cable, able to carry frequencies up to 10 GHz (Suggested Cable). This is the default configuration when using digital demodulation and offsetting the pulse frequency using the AWG.
RCVR I
RCVR Q
Connect the RCVR I and RCVR Q channel to the digitizer or oscilloscope. The EPR signal will appear here.
AWG I
AWG Q
Connect the AWG I and AWG Q connectors to the arbitrary waveform generator. Alternatively, these channels can also be connected directly to the output of a pulse programmer to generate rectangular pulses at the LO1 frequency. If you have questions about these mode, please contact Bridge12 at support@bridge12.com.
PFU TX This is the output of the X-Band IF system. Connect this connector to the input of the high-power pulse amplifier. Depending on the type of the amplifier an additional pulse blanking gate is required as a channel of the pulse programmer.
PFU
BLNK
Blanking gate of the PFU. This is not the amplifier blanking gate. This gate is active high and needs to be high to be able to transmit microwave pulses.
10 MHz
CLK
Connect the system clock to each, the digitizer and the AWG. That way, the X-Band IF system provides the master clock to synchronize all other instruments. The X-Band IF master clock runs at 10 MHz. In cases when the AWG clock is higher, the user may want to synchronize the system to the AWG clock to minimize pulse jitter.

1.1.4 - Manual Operation of the X-Band IF

Required software to install to manually operate the X-Band IF system.

To manually operate the X-Band IF system please install the following software:

Once these two software packages are installed, the X-Band IF system can be completely remote controlled.

1.1.4.1 - Controlling the X-Band IF Synthesizer

How to control the X-Band IF microwave synthesizer manually

Please make sure the control software for the synthesizer is installed. If you haven’t done so, you can download it here.

1.1.4.2 - DAQ Control

How to control the analog and digital lines of the X-Band IF system manually.

Please make sure the control software for the DAQ interface is installed. If you haven’t done so, you can download it here.

Connecting to the DAQ Interface

To connect to the DAQ interface follow these steps:

  1. Make sure the X-Band IF USB interface is connected to the host computer.
  2. Start Kipling
  3. In the GUI click the green USB button (see figure above, left). The USB button will turn red, indicating that Kipling is connected to the DAQ interface (see figure above, right).

Digital Input/Output Control

To control the digital input/output lines click on the menu item Dashboard (see figure above). Each of the lines can be configured as an input or output by selecting the control from the drop-down menu. If the line is configured as an input the indicator to the left will show whether the line is logic high or low. If the line is configured as an output, the status can be changed by selecting the desired status from the drop-down menu.

Below is a list of the different digital channels used by the X-Band IF system.

Pulse Forming Unit (PFU)

Bit Name Function
CIO0 AWG LO select for PFU. 0 - Internal synthesizer (LO1), 1 - AUX LO, auxillary LO input on back panel

Receiver (RCVR)

Bit Name Function
CIO1 Enable/Disable video amplifiers. 0 - disabled, 1 - enabled
FIO4 Microwave signal amplifier, 0 - disabled, 1 - enabled +20 dB gain

Analog Output Control

The value of the various analog outputs can be controlled from the Register Matrix panel. The values for TDAC0 - TDAC3 can be set between -10 and +10 V, with a resolution of 16 bit.

Below is a list of the different analog channels used by the X-Band IF system.

Receiver

Bit Name Function
TDAC0 Video amplifier gain control (I channel), value from -10 to 10 V, -10 V - 30 dB gain, 10 V - 65 dB gain
TDAC1 Video amplifier gain control (Q channel), value from -10 to 10 V, -10 V - 30 dB gain, 10 V - 65 dB gain
TDAC2 LO level receiver, value from -10 to 10 V, -10 V - 0 attenuation, full power on LO of IQ mixer (RCVR), 10 V - 30 dB attenuation, lowest power on LO of IQ mixer (RCVR)

Miscellaneous Other Controls

The X-Band IF system has several spare digital and analog input and output lines. These lines can be configured by the user.

Back Panel EXT Connector (9 pin sub-D)

Bit Name Pin Number
sub-D
Function
EIO0 1 DAQ bit EIO0, can be configured as input and output, TTL level
EIO2 2 DAQ bit EIO2, can be configured as input and output, TTL level
EIO4 3 DAQ bit EIO4, can be configured as input and output, TTL level
EIO6 4 DAQ bit EIO6, can be configured as input and output, TTL level
GND 5 Systems GND
EIO1 6 DAQ bit EIO1, can be configured as input and output, TTL level
EIO3 7 DAQ bit EIO3, can be configured as input and output, TTL level
EIO5 8 DAQ bit EIO5, can be configured as input and output, TTL level
EIO7 9 DAQ bit EIO7, can be configured as input and output, TTL level

Back Panel DIO Connector (9 pin sub-D)

Bit Name Pin Number
sub-D
Function
MIO0 1 DAQ bit MIO 0, can be configured as input and output (TTL level)
MIO2 2 DAQ bit MIO 2, can be configured as input and output (TTL level)
DAC1 3 DAQ bit DAC 1. This is an analog output. The value can be changed between 0 and 5 V. The output is controlled from the Dashboard menu
AIN1 4 DAQ bit EIO 6, can be configured as input and output (TTL level)
GND 5 Systems GND
MIO1 6 DAQ bit MIO 1, can be configured as input and output (TTL level)
IO4 7 TDAC 3, values (set in software) -10 to 10 V, output 0 to 10 V
AIN0 8 DAQ bit EIO 5, can be configured as input and output (TTL level)
VS 9 5 V supply voltage

Back Panel VCA Connector (SMA)

Bit Name Function
DAC0 VCA control (back panel). This is an analog output. The value can be changed between 0 and 5 V. The output is controlled from the Dashboard menu, value 0 to 5 V

1.2 - Frequency Extensions

Frequency extensions for the the X-Band IF System

1.2.1 - Q-Band Extension

Q-Band (34 GHz) Frequency Extension for the Bridge12 X-IF System for EPR Spectroscopy

1.3 - Magnets

Superconducting- and Electromagnets for EPR Spectroscopy

1.3.1 - Superconducting Magnets

Superconducting Magnets for EPR Spectroscopy

Many Bridge12 (high-field) EPR spectrometers use superconducting magnets manufactured by CRYOGENIC LTD. The magnet system comes with extensive documentation.

All information provided here is specific to superconducting magnets specifically manufactured to be used in Bridge12 instruments. In addition, most instructions here describe procedures for day-to-day operations.

1.3.1.1 - Magnet Installation

Supporting Information for Magnet Installation

During the initial installation of the spectrometer, the magnet will be installed by an engineer from Cryogenic Ltd. Once the magnet is installed and passes the customer acceptance test no further installation steps are necessary.

If you have any more questions please reach out to:

1.3.1.2 - Cooling Down the Magnet

How to cool down the magnet

Warning

Please make sure to read the instructions for cooling down the magnet carefully. Make sure you don’t have any questions and all required equipment is in place.

These instructions are not to be used for initial cooling down of the magnet. For these instructions it is assumed that the magnet was properly decommissioned by Bridge12 or CRYOGENIC.

If you have additional questions please do not hesitate to reach out to Bridge12 or CRYOGENIC.

Required Equipment

For cooling down the magnet or recharging the helium reservoir of the integrated VTI (iVTI) the following additional equipment is required:

  • Vacuum pump, preferable a turbomolecular pump with a capacity of 12 m3/hr.
  • Various vacuum hoses, matching center rings and clamps
  • High grade helium gas (grade 5). This is only required for recharging the helium compressor or the helium reservoir of the iVTI. For changing the sample regular grade helium gas is sufficient.
  • Hoses, fittings, pressure regulator, etc. to connect the helium cylinder to the iVTI.

1. Evacuation of the Magnet Cryostat

The magnet cryostat has a single vacuum space which is evacuated via the bellows sealed valve fitted to the cryostat top plate. As a guide, this space should be evacuated using a pump with a capacity of at least 12 m3/hr and a base pressure of <10-4 mbar. Outgassing of the superinsulation occurs whenever the cryostat is allowed to warm up to room temperature. Also, after a magnet quench it is sometimes necessary to evacuate the magnet cryostat, especially if you can’t reach the normal base temperature of the magnet after a quench.

To evacuate the magnet cryostat, connect a vacuum pump to the vale located at the top of the cryostat. Start the vacuum pump and once the pump has reached its base pressure open the valve to the magnet cryostat. The cryostat should be pumped down to a pressure of <10-3 mbar before attempting to cool down. If the pressure decrease is slow, this may be an indication of moisture in the superinsulation which will require more extended pumping of the interspace.

Under normal circumstances there is no need to vent the vacuum space to atmospheric pressure. It is strongly recommended to not vent the vacuum space to decrease the time it takes for the magnet to come up to room temperature. If you have to vent the magnet cryostat please consult the instructions in the magnet manual or contact CRYOGENIC.

2. Prior Checks Before Cooling Down the Magnet

Before cooling down the magnet, please observe the following operational checks:

  • Magnet
    1. The cryostat vacuum space is adequately evacuated to a base pressure of <10e-3 mbar
    2. The compressor is connected to the electrical mains according to the supply information on the compressor housing. Please refer to the compressor manual for further information regarding power requirements.
    3. Adequate water cooling is provided for the compressor (water-cooled versions only). Please refer to the compressor manual for the correct specification for the water circuit.
    4. Both ends of the flexible hoses connecting the compressor to the cold head are properly tightened. Make sure the charcoal trap is installed. Please use the supplied tools to tighten all connections.
    5. The electrical lead from the compressor to the cold head is connected.
    6. The current leads running from the magnet power supply are securely attached to the magnet terminals.
    7. The magnet coil is electrically isolated from the cryostat. To check this, measure the resistance between a magnet current terminal and one of the cryostat top plate outer fixing bolts. A value of ≥ 500 kΩ is typical.
    8. The temperature monitors/controllers are installed properly and the lead(s) are connected.
    9. The static helium pressure in the compressor is correct to ensure it is at the recommended value. Please refer to the compressor manual for the correct specification of the helium pressure. If the pressure is lower than expected please follow the instructions in the compressor manual before starting the compressor.
    10. The system is properly positioned such that there are no significant ferromagnetic objects (e.g.; steel beams, pumping stations etc) close to the cryostat. It should be noted that structural supports on the floor must also be considered when positioning the cryostat. Please contact CRYOGENIC if you are unsure about the possible influence of magnetic objects adjacent to the system.
  • integrated VTI (iVTI)
    1. The dry pump for the iVTI is connected to an outlet. At this time, the dry pump is not powered on.
    2. The external gas reservoir must be charged to the correct pressure with clean helium gas and connected to the cryostat (see system specification). The reservoir is connected to the iVTI gas circuit via a face-sealing connector.
    3. It is recommended that the flow through the iVTI is checked at room temperature prior to cooling down the magnet.
      1. Make sure the dry pump bypass valve (V16) is closed and th helium reservoir valve (V12) is opened.
      2. Make sure the iVTI butterfly valve is open
      3. Switch on the dry pump and pumping for a short time on the iVTI head with the needle valve fully open and circulating gas through the external dump as in normal operation at cold temperatures. The expected flow should be in the range of 5-15 mbar.
      4. Having checked the room temperature flow, switch oof the dry pump
      5. Open the bypass valve (V16). This allows gas to circulate via natural convection throughout the gas circuit during the cooldown of the magnet. The convective flow helps to cool the internal components of the gas circuit whilst minimizing the wear of the pump.
    4. Charge the static column, the space where the resonator is located, to 1 atmosphere of clean helium gas from an external source of helium. The gas is necessary to conduct heat from the wall of the static column and maintain a small temperature differential between the helium flowing in the annular space around static column and the sample probe. It is recommended to charge the static column prior to cooldown as once the system is cold it is more difficult to judge the quantity of helium being added due to condensation of the gas as it enters the column. The approximate volume of the static column is .05 litres.

Cooling Down the Magnet

Once all prior checks have been completed (see instructions above) the magnet cooldown can be started. It is recommended, that the system monitor is started to log all temperatures during the cooldown process.

Cooling the magnet is simply a matter of starting the helium compressor:

  1. Turn the switch located on the back of the compressor into the ON position.
  2. Press the ON button to start the compressor.

Typically the 1st stage of the cryocooler will cool fastest followed by the 2nd stage and then the magnet. For this magnet, the cooldown time is typically about 22 hrs. It should be noted that initially, the operating pressure at the compressor will be high and then reduce (typically by 20 %) as the magnet temperature drops and the cryocooler is required to do less work to remove heat from the system. Once the base temperature of the magnet is achieved the magnet may be energized.

Temperatures of a Typical Magnet Cooldown

Temperatures of a Typical Magnet Cooldown

Temperatures for a typical cooldown of the magnet are shown in the figure above. The following table has shows some typical temperatures once the magnet is at its base temperature.

Q-Band Magnet with iVTI Base Temperatures

Parameter Value Parameter Value
Cryocooler 1st stage 36.2 K Persistence Switch 3.0
Cryocooler 2nd stage 3.0 K Helium pot (Not heated) 3.0 K
Magnet LHS winding 3.0 K Helium pot (heated) 3.2 K
Magnet RHS winding 3.0 K Charcoal trap (not circulating) 46.6 K

Checking the iVTI Operation

Once the magnet has avhieved base temperature:

  1. Close the bypass valve (V16) across the dry pump.
  2. Make sure the cryostat heater is switched off.
  3. Start the dry pump.
  4. Adjust the needle valve to a pressure of approximately 6 mbar at the iVTI pump port. Please note the needle valve may require several further adjustments as the temperature profile within the flow circuit varies whilst steady-state flow conditions are established in the circuit.

For further information/instructions on how to operate the iVTI please see the section about operating the iVTI.

1.3.1.3 - Warming Up the Magnet

How to warm up the magnet

Warning

Please make sure to read the instructions for warming up the magnet carefully. Make sure you don’t have any questions and all required equipment is in place.

If you have additional questions please do not hesitate to reach out to Bridge12 or CRYOGENIC.

Once experiments are finished and the user anticipates some considerable downtown when no experiments are planned the magnet can be warmed up to room temperature. To warm up the magnet:

  1. Make sure the field is at 0 T. This can be verified either in the spectrometer control software or on the front panel of the magnet power supply.
  2. Make sure the spectrometer logger is running and is recording vital data of the system.
  3. To switch off the helium compressor press the off switch located on the back of the device. Once the compressor is switched off, the magnet starts immediately to warm up. It is recommended that the system then be left to attain room temperature naturally which will take approximately 36 - 48 hours.

Warning

Do not disconnect the compressor hoses for any reason during the warm-up of the system. The gas in the cryocooler head must be allowed to expand into the reservoir volume located in the compressor.

It is common practice to break the interspace vacuum when required to warm a cryostat more rapidly than would otherwise occur naturally. CRYOGENIC strongly discourages users from venting the vacuum interspace for any reason. The radiation shield may collapse if subject to an appreciable pressure differential (i.e. by opening the cryostat vacuum valve too quickly).

When the user is ready to continue experiments the magnet needs to be cooled down. Please follow the instructions in the Magnet Cooldown Section.

1.4 - Integrated VTI (iVTI)

1.4.1 - Overview of the iVTI Operation

How the Integrated VTI (iVTI) Works

iVTI - How it Works

The integrated Variable Temperature Insert (iVTI) is a liquid cryogen-free (dry) cryostat built into the superconducting magnet of the EPR spectrometer. It operates by circulating helium gas around a closed-loop circuit cooled by the same cryocooler that is used to cool the superconducting magnet coil. The iVTI can operate continuously and in conjunction with the magnet.

Schematic of the iVTI

Schematic of the iVTI

A schematic of the iVTI is shown above. High quality helium gas (grade 5) is stored at room temperature in the helium reservoir and a dry (oil-free) pump drives drives the circulation of the helium gas through a closed cycle circuit. Under normal operations, the bypass valve (V16) is closed and the helium reservoir valve (V12) and iVTI outlet valve (V11) is open. The helium gas passes from the helium reservoir and into the iVTI circuit via the self-sealing helium gas inlet.

Once the helium gas enters the magnet space, it first passes through a charcoal filter to remove any impurities within the gas. It then flows through the first stage heat exchanger which cools the gas to ~40K. The gas then passes to the second stage heat exchanger where it is cooled further to below 4.2K and condenses in the helium pot. The helium pot has an integrated heater to keep the pot at a recommended temperature of 3.1 K.

To cool the cryostat space, the liquid helium then flows across the needle valve, at which point it expands and cools further to approximately 1.6K. The cooled gas/liquid mixture then passes through the iVTI heat exchanger where the helium is heated to the required temperature. The helium gas flows upwards in an annular space surrounding the static column (containing the EPR probe) to the top of the iVTI where it exits and is circulated through the pump and exhausts back to the dump vessel.

The probe is cooled using an exchange gas, which is cooled at the wall of the static column.

The temperature inside the cryostat is regulated using the iVTI heater located at the bottom of the static column, and the heating power is regulated by the temperature controller. To achieve the most accurate temperature at the position of the sample, all Bridge12 EPR probes are equipped with a calibrated Cernox temperature sensor. This sensor is located as close to the sample as possible for accurate temperature readings and this temperature is used to control the heater power in a feedback loop.

If the probe is removed from the cryostat, the temperature can be regulated using the internal iVTI temperature sensor. Note, that the iVTI temperature is not necessarily equal to the sample temperature.

1.4.2 - Getting Started with the iVTI

Getting Started with the iVTI

Important

Before operating the iVTI please familiarize yourself with the principle of operations described in the section iVTI Overview.

The following section provides information about day-to-day operation of the cryostat. If you need assistance troubleshooting iVTI operations, please first consult the manual or contact Bridge12 or CRYOGENIC.

Prior Checks Before Operating the iVTI

Warning

Do not use excessive force or overtighten the needle valve as this can lead to permanent damages of the valve.

Note

If the needle valve is closed and open again, a faint clicking sound is heard as the needle pulls away from the needle valve seat.

Once the magnet has achieved its base temperature the following checks should be performed prior to operating the iVTI:

  1. Make sure the dry pump bypass valve (V16) is closed.
  2. Make sure the helium reservoir valve (V12) and the iVTI outlet valve (V11) are open.
  3. Start the dry pump and adjust the needle valve to a pressure of approximately 5-8 mbar at the head of the iVTI. Please note the needle valve may require several further adjustments as the temperature profile within the flow circuit varies whilst steady-state flow conditions are established in the circuit.

Within a short period of time, the cryostat should reach its base temperature of ~ 1.8 K.

Setting an optimum flow rate requires some experience from the user. The factors that must be considered are:

  • If the flow rate is too low the 2 K base temperature of the iVTI may not be achievable, as the available cooling power of the gas may be too low.
  • If the flow rate is too high the cryocooler second stage temperature will rise which may affect the maximum ramp rate of the magnet.
  • If the flow rate is too high the cryocooler may not condense at a fast enough rate. This leads to rapid boiling of the condensed gas in the circuit and a rapid pressure rise of the circuit.

Typically the iVTI can operate with a single needle valve setting over its whole temperature range. However, if a rapid cooldown is required the user can increase the flow and then subsequently re-adjust the needle valve to give a lower flow setting once the iVTI has cooled close to the required temperature.

1.4.3 - Operating the iVTI

How to Operate the iVTI

Important

Before operating the iVTI please familiarize yourself with the principle of operations described in the section iVTI Overview.

The following section provides information about day-to-day operation of the cryostat. If you need assistance troubleshooting iVTI operations, please first consult the manual or contact Bridge12 or CRYOGENIC.

Schematic of the iVTI

Schematic of the iVTI

The iVTI operates by circulating helium gas in a closed-loop circuit. A schematic of the circuit is shown above. The following instructions assume the magnet is at its base temperature and the iVTI has passed operational checks.

Cooldown of the iVTI

To cooldown the cryostat follow these instructions:

  1. Make sure the dry pump bypass valve (V16) is closed.
  2. Make sure the helium reservoir valve (V12) and the iVTI outlet valve (V11) are open.
  3. Start the dry pump and adjust the needle valve to a pressure of approximately 5-8 mbar at the head of the iVTI. Please note the needle valve may require several further adjustments as the temperature profile within the flow circuit varies whilst steady-state flow conditions are established in the circuit.

While the cryostat cools down, the pressure of the helium reservoir will decrease as more helium condenses in the helium pot.

Setting the Cryostat Temperature and Regulating the Helium Flow

Setting the Temperature

The temperature at the position of the sample is set at the temperature controller. This can be done either manually at the front panel of the temperature controller or through the spectrometer control software SpecMan4EPR. Under normal operations, when there is an EPR probe installed, the temperature of the cryostat is regulated based on the probe temperature sensor readings to give the most accurate control over the temperature at the position of the sample. Once the temperature is set, either on the front panel or through SpecMan4EPR, the user needs to adjust the helium flow by adjusting the needle valve. However, once the desired temperature is reached, no further optimization is required.

Optimization of the temperature controller PID and heater range parameters is beneficial if the ultimate system performance is to be achieved. The parameters may vary depending upon the flow rate established in the circuit. Typical starting values for the PID settings are 250,120,30 which will provide reasonable control over the whole temperature range. However, the user is encouraged to adjust PID settings to further optimize the system as an experience of the iVTI is established. The heat exchanger is fitted with a heater (30 Ohm / 30 W). Under no circumstances should a current greater than 1A be applied to the heat exchanger heater, as it will fuse the leads.

Setting the Needle Valve for Optimum Flow Control

The needle valve creates a controllable impedance for the helium circulation. It is set initially and thereafter should need only occasional adjustment. A dial gauge is fitted at the top of the iVTI to allow the user to set and monitor the flow rate (directly related to pressure). A pressure between 5 and 15 mbar at the top of the iVTI is recommended. The optimum value is specific to each system.

Under normal conditions, a pressure of about 10 mbar can cover a wide range of temperatures. Especially, for experiments that are performed at about 50 to 60 K, the position of the needle valve rarely has to be changed.

To operate at temperatures below 10 K, the pressure may have to be adjusted using the needle valve. In this case, the optimum value will correspond to the vapour pressure curve as shown in the figure below.

Helium Vapor Pressure

Helium Vapor Pressure

In general, if the helium flow rate is too low there may not be enough cooling power available from the circulating helium and the system may not reach the base temperature of the iVTI. A slightly too high a flow rate and the system will not reach base temperature because it is not following the vapour pressure curve.

However, if the flow rate is far too high the amount of heat that needs to be extracted from the circulating helium will exceed the cooling power of the cryocooler. In this case, the temperature of the cryocooler 2nd stage, magnet and helium pot will increase and the liquefaction of helium in the iVTI circuit will slow or stop. Under these circumstances, the helium pot will eventually empty preventing the iVTI from achieving base temperature. The associated increase in temperature will also affect the magnet ramp rates.

It should be noted that the helium flow is also affected by the impedance of the tubes close to the iVTI heat exchanger. The overall impedance depends on the temperature of the heat exchanger. A higher driving pressure is needed between the helium pot and the iVTI when the iVTI is at a high temperature than when it is at a low temperature. The liquid helium in the helium pot is effectively in equilibrium. So the pressure of the gas above the liquid in the helium pot follows the vapour pressure curve for helium 4 shown in the figure above. Since the inlet circuit has a low impedance, the pressure in the helium reservoir is close to that at the helium pot and can be used as a gauge as to how much helium has been condensed into the pot. A heater on the helium pot is used to control the temperature of the helium pot and hence control the driving pressure. The temperature is set to a recommended value of 3.1 K. The helium pot temperature is controlled via a software PID loop.

Changing the Sample

During the process of changing the sample, the iVTI circuit will continue to run and the dry pump should not be switched off. To change the sample, follow these instructions:

  • Removing the Sample
    1. If you want to insert/replace a new sample, make sure the sample is properly mounted on the sample stick. If you want to remove the sample without inserting a new sample, make sure you have a blind plug handy.
    2. Make sure the experiment is finished.
    3. Make sure helium purge gas is available and the pressure regulator is set to the correct value. For the sample exchange circuit, the pressure regulator should be set to a pressure of 3-5 psi.
    4. Open the valve to the cryostat, located at the top of the magnet and observe the pressure inside the cryostat. You should charge the cryostat space to about 1 psi. A safety valve, located close to the valve will release all excess pressure.
    5. Loosen the nut securing the sample stick.
    6. Slowly pull out the sample stick. The sample stick has holes at the top and bottom of the G10 section (green). Once the top hole passes the nut securing the sample stick, cold helium gas will exit the hole. Pull the sample stick out of the probe and properly store the sample. While the purge gas is running, air is prevented to enter the sample space.
  • Inserting the Sample
    1. With the purge gas running, insert the sample stick into the probe.
    2. Slowly lower the sample stick into the probe. Once the bottom venting holes of the sample stick have past the nut securing the sample stick the helium purge gas will exit at the top holes of the sample stick. Wait for a couple seconds to purge the space inside the of the sample stick.
      • Optionally, open the nut at the top of the sample stick to purge the space inside the stainless steel section of the sample stick.
    3. Slowly push the sample stick into the resonator. At this point, the user should also monitor the resonator tuning in the control software. Once the sample tube is entering the resonator, the frequency will shift to a lower value. Push the sample stick until it is properly seated at the top of the resonator.
    4. Tighten the nut securing the sample stick. Important, this only needs to be hand-tightened. Do not use any tools to tighten this nut.
    5. Close the valve to the helium purge gas supply.
    6. Switch on the diaphragm pump to evacuate the sample space. Lower the pressure to about 200 - 500 mbar. The exact value depends on desired temperature and will require some user experience.
    7. Close the valve to the iVTI located at the top of the magnet.
    8. Turn off the diaphragm pump.

Once the sample is loaded into the resonator make sure to close all valves to the helium supply for purging.

1.4.4 - Troubleshooting iVTI Operation

How to Operate the iVTI

Important

Before operating the iVTI please familiarize yourself with the principle of operations described in the section iVTI Overview.

The following section provides information about day-to-day operation of the cryostat. If you need assistance troubleshooting iVTI operations, please first consult the manual or contact Bridge12 or CRYOGENIC.

Blockage in the iVTI Circuit

The most likely cause of abnormal iVTI performance is a blockage in the iVTI circuit due to ingress of air.

Symptoms of a iVTI partial blockage before the helium pot can include:

  • Reduced flow at a given needle valve setting.
  • Reduced pressure at the iVTI pumping side for a given needle valve aperture.
  • Increased pressure at the external helium reservoir.
  • Decreased temperature of the 2nd stage and helium pot.

The pressure in the helium reservoir increases as the partial blockage limits the flow of helium through the inlet heat exchangers. The 2nd stage and helium pot temperatures may fall as the cryocooler heat load is reduced due to the lower helium flow rate at the 2nd stage.

Symptoms of a blockage after the helium pot are similar; however, in this case, the helium reservoir pressure may fall as the helium charge is condensed into the system with a reduced flow returning to the reservoir.

If a blockage is suspected the following actions are recommended:

  • Heat the iVTI to room temperature. In some cases, this may clear the blockage.

If the blockage persists, please follow these steps:

  1. Once the iVTI is at room temperature, disconnect the helium reservoir hose at the face-sealing connector on the cryostat.
  2. Connect an external pump to the face-sealing connector of the cryostat using the fitting that was provided with the system. This connection provides the most efficient pumping of the inlet circuit which is the most likely location for any blockage to occur.
  3. Close the valve at the head of the iVTI (V11) to the dry pump and open the needle valve fully.
  4. Pump the iVTI via the external pump and switch off the compressor.
  5. Heat the charcoal trap using the temperature controller and maintain it at 300 K.
  6. Wait for the 2nd stage to rise in temperature above 90 K.
  7. Turn off the charcoal heater, isolate the external pump and restart normal flow through the iVTI circuit.

If the blockage persists, continue to warm the system to room temperature and replace the helium of the iVTI circuit as outlined below.

Replacing the Helium in the iVTI Circuit

If contamination or loss of helium in the iVTI circuit is suspected please follow these instructions to replace the helium charge:

  1. Make sure the system is at room temperature and the dry pump is not running.
  2. Connect a T-connector to the plug valve on the exhaust side of the dry pump. One arm of the T should be connected to a clean helium source with an in-line valve, and the other to a pump via a valve. The helium source should be either a bottle of high purity gas (99.95%, grade 5) or the boil-off gas from a helium storage vessel. Do not use standard (balloon) grade bottled helium for this purpose as this can have a high moisture content.
  3. Open the connection to the external pump and pump out the helium reservoir. Once the pressure is low enough (< 1e-3 mbar>) activate the dry pump and use it to purge the internal iVTI circuit.
  4. Close the valve to the external pump and fill the reservoir to a pressure of +1psi above atmospheric pressure with pure helium gas.

The system is now ready for cooldown and operation.

1.5 - The Bridge12 Q-Band Spectrometer

Online documentation for the Bridge12 Q-Band pulsed EPR spectrometer.

1.5.1 - Overview

Compact Q-Band EPR Spectrometer

The Bridge12 Q-Band spectrometer is a compact EPR spectrometer for pulsed EPR spectroscopy.

Bridge12 Compact Q-Band EPR Spectrometer

The main features of the system are:

For more information about the system visit the Q-Band Spectrometer Page or reach out at info@bridge12.com.

1.5.2 - Getting Started

Getting started with the Q-Band spectrometer SpecMan4EPR

In this section we give a general overview how to operate the Bridge12 Q-Band pulsed EPR spectrometer, which will include the following steps:

  • Loading the sample into the spectrometer
  • Optimizing the microwave pulse parameters
  • Record an echo-detected field sweep EPR experiment
  • Perform a PELDOR/DEER experiment

All experiments were performed on our in-house spectrometer, using a sample of a ruler molecule with a concentration of 100 µM in polystyrene. Experiments were performed at a temperature of 50 K.

Important

These are sample instructions recorded on equipment in our lab. Actual parameters may vary depending on your specific system configuration, sample, or sample temperature.

This documentation was prepared using SpecMan4EPR version 3.6.4. If you use a different version some features are maybe named differently or the layout of some windows has changed.

Preparing SpecMan4EPR

When you first start SpecMan4EPR, the software will make sure it can connect to all devices. Once the initialization routine finishes you will see a window very similar to the one shown in the figure below.

SpecMan4EPR

We recommend to open the following windows to operate the spectrometer:

  1. Open the Scope Window and activate the digitizer trace. This will be helpful to see the integration limits for the digitizier.
  2. Open the Pulse Programmer window to show the pulse sequence.
  3. Open the top User Panel, this will give you convenient access to bridge parameters such as the microwave frequency, video amplifier gain, or the microwave phase.
  4. Optionally, open the Log Window. This is helpful to see messages from the SpecMan4EPR software.

Once you opened all these additional panels, the SpecMan4EPR window should look similar to the one shown below.

SpecMan4EPR with additional panels

What’s next?

1.5.2.1 - Inserting a Sample and Adjust the Resonator Coupling

How to insert the sample into the spectrometer

The following section provides a step-by-step procedure how to insert the sample into the EPR probe and adjust the iris coupling of the resonator.

To confirm that the sample is inserted into the microwave resonator correctly, we will observe the microwave tune dip of the resonator.

Inserting the Sample

Make sure SpecMan4EPR is running.

To insert the sample follow these steps:

  1. Switch the spectrometer into operate mode
    1. From the Experiment Panel select the mode tab
    2. Click the Operate button. This will enable all synthesizers
SpecMan4EPR mode panel
  1. Open the resonator tune window (RESONATOR plugin) from the plugins window
  2. Start the tune sweep by clicking the run button (green arrow on bottom left)
SpecMan4EPR tune mode
  1. Make sure the resonator is critically coupled. For this, the resonator iris is completely inserted into the resonator. This corresponds to a micrometer reading of 0 mm.

To insert the sample:

  1. Make sure the helium gas pressure inside the cryostat is slightly above atmospheric pressure (> 0.1 bar). Note, the pressure gauge shows a value of 0 bar, when the pressure inside the cryostat is close to atmospheric pressure.
  2. Once the pressure in the cryostat is slightly above atmospheric pressure, it is safe to remove the bling plug. For this loosen the nut and pull out the plug. You will feel a stream of cold helium gas exciting the cryostat. This will prevent air and moisture entering the sample space.
  3. Slowly insert the the sample stick and lower it into the resonator. The sample stick has holes at the top and the bottom to release the helium gas coming from the probe. While lowering the sample stick, observe the tuning picture of the resonator. Once the sample enters the resonator, the tuning dip of the resonator will be visible (see figure below).
Resonator tuning with and without sample
  1. Once the sample is fully inserted into the resonator, tighten the nut at the top of the probe that holds the sample stick. This nut should just be hand-tight. Do not use any tools (e.g. pliers) to close the nut.
  2. Close the valve to the helium gas supply.
  3. Start the pump and open the valve to the pump to lower the pressure inside the cryostat to about -0.5 bar. The exact pressure is not important.
  4. Close the valve to the pump and wait for a couple minutes until the temperature inside the cryostat stabilizes.

In side the tune window you can zoom into the tune picture by clicking the up and down buttons.

SpecMan4EPR tune mode

Overcoupling the Resonator

For pulsed EPR experiments the resonator must be overcoupled to reach the maximum resonator bandwidth.

  1. Before you adjust the resonator iris to overcouple the resonator, make sure the microwave frequency (red vertical line) is set to the tuning dip. The microwave frequency is adjusted in the Control Panel (see figure above).
  2. Slowly turn the micrometer screw to overcouple the resonator. This will raise the iris and the tuning dip will get broader. The tune dip will also shift ub by about 90 MHz. For maximum overcoupling set the height of the iris to 7 mm.
  3. For DEER experiments on nitroxide-based spin labels you don’t have to adjust the microwave frequency. The shift of the dip is about 90 MHz, which corresponds to the frequency separation of the pump and observe pulse. When using different paramagnetic species the frequency may have to be adjusted. However, keep in mind, the resonator bandwidth is > 400 MHz for the Bridge12 QLP probe.
  4. To exit the tune mode, click the stop button (click the square icon, bottom left) and click the OK button to close the window.

The spectrometer is ready for the first experiment.

What’s next?

1.5.2.2 - Optimizing the Microwave Pulse Parameters

How to optimize the microwave pulse parameters for PELDOR/DEER experiments

In this section we provide the general procedure how to optimize the microwave pulse parameters for the PELDOR/DEER experiment.

Set the Magnetic Field

  1. Use the microwave frequency that was set in the microwave bridge Control Panel to calculate the maximum field for a nitroxide radical. This can be done using the provided Python script.
    • In this example the microwave frequency is 34.032 GHz and the script returns a field value of 1.21896 T.
  2. Next, open the window to control the field. From the DDI menu select Set Field form, FLD.
  3. Enter the field value and click the Set Field button.
SpecMan4EPR magnetic field control
  1. Wait until the desired field value is reached. When the magnetic field is sweeping, the temperature of the Magnet Switch will typically rise from about 5.9 K to about 8 K. Once the desired field is reached, the temperature returns to its base temperature of about 5.9 K.
  2. Close the magnet control window.

Optimizing the Pulse Parameters

To optimize the pulse parameters follow these steps:

  1. Load a new experiment. From the File menu, select New Experiment. Scroll to the Two-Pulse Echo experiment and click on it to load the experiment. This will load some default pulse parameters, which will be a good start for the optimization.
  2. Next, start the pulse sequence by hitting the Tune button. Once the experiment starts, you should see an echo in the Scope window to the left.

In the next step we will adjust the video amplifier gain and the microwave phase. This is an iterative process and you have to adjust both parameters. Make sure the Modulation Frequency parameter is set to 500 MHz.

  1. In the microwave bridge Control Panel adjust the video amplifier gain. Observe the echo amplitude and set the video amplifier gain to a value so the echo amplitude is not clipped.
  2. In the same window adjust the microwave phase to maximize the echo amplitude (real part of the complex signal, green trace).

The video amplifier gain should be adjusted to a value that the echo amplitude is about 75 % of the maximum (horizontal red line). For small echo amplitudes, you can also adjust the scale in the same window. This will adjust the input scale of the digitizer.

In the next step, we will find the optimum value for the microwave power. SpecMan4EPR allows the user to make almost every variable of the experiment an axis of the experiment. We will use this feature of the software to find the correct pulse amplitude.

  1. Right click on the amp variable in the Parameter section and select Y axis to assign the parameter to the y axis.

  2. In the section for the y axis, set the size of the axis to 101 points.

  3. In the line below, set the sweep range to "0 step 0.01". This will you to step the amp parameter from 0 to 1 in steps of 0.01 using the slider.

  4. To start the experiment, click the tune button.

  5. Move the slider with the mouse to maximize the echo amplitude.

    1. Find the optimum pulse amplitude for a two-pulse echo using 8 and 16 ns for the pi/2 and pi pulse, respectively. In this example the optimum value is at 0.2 (21st step).
    2. Change the pulse length the for 90º and 180º pulse to 16 and 32 ns, respectively, and repeat the optimization. In this example the optimum value for the 16/32 echo is at 0.08. This value should be approximately half of the value determined for the 8/16 ns echo.

    While the tune mode is running, SpecMan4EPR will integrate the area of the echo and record the value in the display. Observe the value (red trace in figure below) and optimize for the maximum echo amplitude.

  6. Once you found the optimum values stop the tune mode by clicking the tune button.

SpecMan4EPR optimizing pulse parameters

Record reference T2 decay

In the next step, we will record a two-pulse echo decay as a reference spectrum. This will give you and idea of the setting for the maximum dipolar evolution time. To record the T2 decay follow these steps:

  1. Reload the experiment. From the File menu, select New Experiment. Scroll to the Two-Pulse Echo experiment and click on it to load the experiment.
  2. Set the amp parameter (pulse amplitude) to the value determined above. In this example we use the parameters for the 8/16 ns echo.
  3. Hit the run button to start the experiment.

It will take about 15 s to compile the pulse sequency and upload it to the AWG. Once the sequence is uploaded the experiment will start and the results are shown once the experiment finishes (see figure below). Overall duration for this experiment is about 26 s.

SpecMan4EPR two-pulse echo decay

For this particular sample we pick a maximum dipolar evolution time of about 2 µs.

What’s next?

1.5.2.3 - Record a Field Sweep Spectrum

How to record an echo-detected field sweep spectrum

In this section we will provide general instructions on how to record an echo-detected field sweep spectrum.

  1. First, load a new experiment. From the File menu, select New Experiment. Scroll to the Field Sweep Echo in Set Mode experiment and click on it to load the experiment.
  2. When you load the experiment SpecMan4EPR will set all experimental parameters to the default values loaded from the template. Adjust the parameters to the values determined in the previous experiments, such as pulse length, pulse amplitude, …
  3. Under X-axis set the field range and number of points for the experiment. To get started use the field sweep range, that is calculated by the Python script. In this example we will sweep the field from 1.212 T to 1.232 T. enter the phrase “1.212. to 1.232” under Field. The step size is automatically calculated.
  4. Hit the run button to start the experiment.
SpecMan4EPR field sweep spectrum

Once the experiment is finished, data will be saved automatically by SpecMan4EPR. In this particular experiment the software will set a magnetic field and will wait for the field to be settled (SetMode: SetAndWait). While this is the slowest mode to record a spectrum, it will give the most accurate results. At the end of the experiment SpecMan4EPR will return the field to the start value.

Note down the value of the magnetic field for the maximum echo amplitude. You can enable the cursor by enabling the Hint mode:

  1. Click the right button in the top right corner of the spectrum (see figure below)
  2. From the drop-down menu select Show Hint.
SpecMan4EPR hint mode

In this particular experiment the field position for the maximum echo amplitude is 1.21865 T.

For nitroxide spin-labels at Q-Band frequencies the separation between the pump and the observe pulse is typically set to 90 MHz, corresponding to 90 MHz / 2.804 = 3.2097 mT.

The field position for the observe pulse is therefore 1.21865 T + 3.2097 mT = 1.221859 T.

What’s next?

1.5.2.4 - Run a B1 Nutation Experiment

How to run a B1 nutation experiment

In this section we will provide general instructions on how to run a nutation experiment to determine the optimum length of the inversion pulse.

The nutation is a three pulse experiment. The sequence starts with a nutation pulse followed by a long delay and a two-pulse echo detection sequence. During the experiment the pulse length of the nutation pulse is increased to observe the Rabi oscillation.

To perform the nutation experiment follow these steps:

  1. Start by loading the experiment. From the File menu, select New Experiment. Scroll to the Two Pulse Echo Nutation experiment and click on it to load the experiment.
  2. Make sure the magnetic field is still set to a value corresponding to the maximum echo amplitude (here 1.21865 T).
  3. In the experimental parameter section set the parameter amp_Nut to 1. This is the amplitude for the nutation pulse and we will just use maximum power.
  4. Set pulse length to p_90 = 16 ns and p_180 = 32 ns, set the amplitude for the detection pulses to 0.08 (this is the value that was previously determined).
  5. Hit the tune button and adjust the integration interval in the scope window (vertical orange lines).
  6. To start the experiment, hit the run button.

The result of the experiment is shown in the figure below.

SpecMan4EPR Rabi oscillations

Use the cursor and Hint Mode to determine the pulse length of the inversion pulse. In this case it is 10 ns for an pulse amplitude of 1.

Note

The Bridge12 QLP resonator has a bandwidth of > 400 MHz. Therefore, you don’t have to repeat the nutation experiment at the observe field position.

What’s next?

1.5.2.5 - Run a PELDOR/DEER Experiment

How to perform a PELDOR/DEER measurement

If you followed the instructions of the previous sections you have all parameters to run a PELDOR/DEER experiment. In this section we show how to setup a PELDOR/DEER experiment. Note, that some parameters such as shots, trace, scans … are set to default values. We recommend to start with these parameters and adjust them later if needed.

First, we need to move the magnetic field to the observe field position:

  1. Open the window to control the field. From the DDI menu select Set Field form, FLD.
  2. Enter the field value for the observe field (here 1.2218 T) and click the Set Field button.
  3. Wait for the field to settle.

In the next step we can setup the PELDOR/DEER experiment:

  1. Start by loading the experiment. From the File menu, select New Experiment. Scroll to the 4-Pulse DEER experiment and click on it to load the experiment.
  2. Enter the experimental parameters corresponding to the values determined in the previous section:
    1. Set the pulse length for the observe pulses (t90, t180), the 16 and 32 ns, respectively.
    2. Set the amplitude for the observe pulses (amp) to 0.08. This value was determined in the Optimizing Pulse Parameters section.
    3. Set the value for the maximum dipolar evolution time (tau2) to a value of 1.932 µs.
    4. Set the length of the pump pulse (t_eldor) to 10 ns. This value was determined in the Run a B1 Nutation Experiment section.
    5. Set the amplitude of the pump pulse (amp_eldor) to a value of 1 (maximum power).
    6. Set the frequency of the pump pulse (f_ELDOR) to a value of 590 MHz. The detection is at 500 MHz. Therefore, the overall offset of the pump pulse, with respect to the observe pulse is 90 MHz.
  3. Set the parameter t, the position of the pump pulse to “-100 ns step 16 ns” and the size to 126. Here you should pay attention to the value in the bracket. SpecMan4EPR calculates this value and it should be lower than the value for tau2, the maximum dipolar evolution time.
  4. Enter the number of scans that you would like to run.
  5. Hit the run button to start the experiment.

When you hit the run button, SpecMan4EPR first calculates the entire pulse sequence, including the phase cycle (SpecMan4EPR will step through all variables and you can see this for the different axis). The full pulse sequence is then uploaded to the AWG and the experiment is started. Once a scan is finished and the data is transferred to SpecMan4EPR.

SpecMan4EPR PELDOR/DEER trace

Once the experiment is finished, SpecMan4EPR will automatically save the results. If you want to stop the experiment during the acquisition, you have two choices:

  1. Hit the Run button. This will immediately stop the experiment. In this case the data is not saved automatically.
  2. Hit the Finish button. SpecMan4EPR will finish the current scan before stopping the experiment. The experimental data will be saved.

Setup Overnight experiment

If long signal averaging is required to obtain a sufficient signal to noise ratio, we recommend running the experiment in a queue.

  1. In the experiment panel, select the queue tab (see figure below) top open the queue window.
SpecMan4EPR experiment queue
  1. Click + button (with the green + sign) to add an experiment to queue. This will queue an experiment using the parameters that are currently set in the parameter section.
  2. Click the loop button. By selecting this option, SpecMan4EPR will perform the experiments in the queue (here only ne experiment is added to the queue) and then start over again once all experiments in the queue are finished. The experiment is saved once it finishes.
  3. To start the queue, press the queue button.

What’s next?

1.5.2.6 - Shutting Down the Spectrometer

How to shut down the spectrometer once the experiment is finished

Once all experiments have finished you can shut down the spectrometer.

Shutting down the spectrometer means:

  • Closing SpecMan4EPR (if desired)
  • Switching of the microwave amplifier and Q-Band extension
  • Ramping down the magnetic field to 0 T

Shutting down the spectrometer DOES NOT MEAN:

  • Switching off the helium compressor

The helium compressor should only be switched off for maintenance or if the spectrometer is not used for a longer time.

Removing the Sample

Follow these steps to remove the sample:

  1. Critically couple the resonator. This will make it easier to find the resonator the next time. Turn the micrometer screw to 0 mm.
  2. Open the resonator tune window (RESONATOR plugin) from the plugins window
  3. Start the tune sweep by clicking the run button (green arrow on bottom left)

You should see a clean tuning picture (see figure below)

SpecMan4EPR tune mode

To remove the sample:

  1. Make sure you have the blind plug within reach.
  2. Make sure the helium gas pressure inside the cryostat is slightly above atmospheric pressure (> 0.1 bar). Note, the pressure gauge shows a value of 0 bar, when the pressure inside the cryostat is close to atmospheric pressure.
  3. Once the pressure in the cryostat is slightly above atmospheric pressure, it is safe to remove the sample stick. For this loosen the nut and slowly pull out sample stick. While raising the sample stick, observe the tuning picture of the resonator. Once the sample is pulled out of the resonator, the tuning dip of the resonator will disappear (shifted to higher frequency).
  4. Slowly keep raising the sample stick. The sample stick has venting holes at the top and bottom of the G10 section to release helium gas from the cryostat and to purge the sample stick and to prevent air and moisture to enter the sample space.
  5. Once the sample stick is removed from the cryostat secure the sample and either insert a new sample or close the cryostat using the blind plug and tighten the nut. This nut should just be hand-tight. Do not use any tools (e.g. pliers) to close the nut.
  6. Once the cryostat is closed, close the valve to the helium supply.
  7. Start the pump and open the valve to the pump to lower the pressure inside the cryostat to about -0.5 bar. The exact pressure is not important.
  8. Close the valve to the pump.

Ramping Down the Magnetic Field

Once the sample is removed the magnetic field can be ramped down.

  1. Open the window to control the field. From the DDI menu select Set Field form, FLD.
SpecMan4EPR magnetic field control
  1. Enter 0 in the field for the magnetic field and hit return.
  2. Under Sweep rate, select the highest sweep rate.
  3. Click the Set Field button to ramp down the magnetic field.
  4. Wait for the field to reach 0 T.

Switching the Spectrometer to Standby Mode

To switch the spectrometer into operate mode:

  1. Select the mode tab from the Experiment Panel.
  2. Click the Standby button. This will disable all synthesizers.

The spectrometer is in standby mode. It is safe to switch off the amplifier and the Q-Band extension.

2 - Quasi-Optics

Quasi-Optical Systems and Components for EPR/DNP Spectroscopy

Quasi optical (QO) systems are used in high-field EPR and DNP systems to control the microwave beam propagation. For example, to separate the incident microwave beam from the reflected beam in a reflection-type EPR bridge. QO is a highly versatile technology to build attenuators, universal polarizers, and beam splitter. If you would like to learn more about it check out our page on Quasi-Optics (QO).

Here you will find the online documentation for some of our QO components. Many QO systems are customer specific. If you have questions about your individual system please contact Bridge12 at support@bridge12.com.

2.1 - Polarizing Transforming Reflector (PTR)

The Bridge12 Polarization Transforming Reflector (PTR) is a universal polarizer. The PTR can convert a linearly polarized microwave beam into a circularly polarized beam, or rotate the E-field vector by any angle.

Bridge12 Polratization Transforming Reflector (PTR)

Bridge12 Polratization Transforming Reflector (PTR)

This is the online documentation of the PTR. For product related information check out the Bridge12 PTR Product Page. The theory and operation of the PTR has been published in:

  • Chen, Jeson, and Thorsten Maly. “Compact, Tunable Polarization Transforming Reflector for Quasi-Optical Devices Used in Terahertz Science.” Review of Scientific Instruments 93, no. 1 (January 1, 2022): 013102. https://doi.org/10.1063/5.0036292.

If you are using a Bridge12 PTR in your research please consider citing the device in your publication.

2.1.1 - Theory of Operation

The PTR was first described in the literature by Howard et al. operating at a frequency of 28 GHz. Despite its low insertion loss, the PTR has only been used in QO systens for high-field EPR/DNP spectroscopy in some rare instances.

Theory

Wire Grids

Before discussing the PTR in more detail, we will briefly review some fundamental properties of wire grid polarizers.

Free-standing wire grid polarizers made from cylindrical metallic wires either transmit or reflect electromagnetic radiation depending on the orientation (polarization) of the incident E field vector with respect to the grid wire orientation. For cylindrical wires with wire radius \(d\) and wire separation distance \(S\) with \(d \ll S\), the frequency dependent transmission coefficient for an electric field vector oriented parallel to the wire grid can be calculated using the Jones Matrix formalism. The corresponding reflection and transmission matrices are:

$$ C_{R} = \begin{pmatrix} -\cos^{2}{\theta} & -\cos{\theta} \sin{\theta} \\ \cos{\theta} \sin{\theta} & \sin^{2}{\theta} \end{pmatrix} $$

and

$$ C_{T} = \begin{pmatrix} \sin^{2}{\theta} & -\cos{\theta} \sin{\theta} \\ -\cos{\theta} \sin{\theta} & \cos^{2}{\theta} \end{pmatrix}. $$

This only correct if the grid orientation to be perpendicular to the propagation direction of the beam (\(\theta^{ \prime} = 0^{\circ}\)). However, if the purpose of the grid is to clean up the polarization of the microwave beam, the resulting reflected microwave power should be directed away form the source and not directly reflected back. This can be achieved by orienting the grid at an angle of \(\theta^{\prime} = 45^{\circ}\). In this case determining the reflection and transmission coefficients is less intuitive. This is summarized in the following figure (Figure 1).

Figure 1: Grid orientation for partial beam reflection/transmission

Figure 1: Grid orientation for partial beam reflection/transmission

If the grid is oriented perpendicular to the beam propagation direction (\(\theta^{ \prime} = 0^{\circ}\)) the free-standing wire grid needs to be rotated by (\(\theta = 45^{\circ}\)) to transmit 50 % of the power (see Figure 1A). On the other hand, if the grid is oriented at an angle (\(\theta^{ \prime} = 45^{\circ}\)) the wire grid needs to be rotated by (\(\theta = 54.7^{\circ}\)) to transmit 50 % of the power and reflect 50 % of the power (see Figure 1B). For more details check out Paul Goldsmith’s book Quasioptical Systems: Gaussian Beam Quasioptical Propagation and Applications (chapter 8.6.2, page 210).

Figure 2: The free-standing wire grid as a beam splitter

Figure 2: The free-standing wire grid as a beam splitter

The power transmission and reflection for a variable grid is shown in Figure 2. By rotating the grid and varying the angle \(\theta\) from 0º to 90º the beam can be continuously attenuated. Depending on the quality of the grid this attenuation can reach values of - 30 dB or lower.

Polarization Transforming Reflector (PTR)

The PTR consists of two principle components: 1) A free-standing wire grid and 2) A flat metal reflector. The free-standing wire grid is made from metallic wires having a diameter of d and separated by a distance of S (see figure below). Directly located behind the grid is a flat mirror. A schematic of the PTR is shown in Figure 2 (left).

Figure 2: Bridge12 Polarization Transforming Reflector (PTR). Left: Schematic of the PTR. Right: Dependance of the output polarization from the mirror/grid separation distance. Ev - vertically polarized e-field, Eh - horizontally polarized e-field, Er - right hand circularly polarized e-field, El - left hand circularly polarized e-field.

Figure 2: Bridge12 Polarization Transforming Reflector (PTR). Left: Schematic of the PTR. Right: Dependance of the output polarization from the mirror/grid separation distance. Ev - vertically polarized e-field, Eh - horizontally polarized e-field, Er - right hand circularly polarized e-field, El - left hand circularly polarized e-field.

By changing t, the distance between the flat mirror and the wire grid polarizer of the PTR it is possible to change the polarization state of e.g. a linearly polarized Gaussian incident beam to from linear to circular clockwise and counter-clockwise polarization (Figure 2, right).

How it works

Assuming the gird/mirror pair of the PTR is oriented at angle \(\theta = 45^{\circ}\) with respect to the propagation direction of the incident beam and a grid wire orientation of (\(\theta^{ \prime} = 54.7^{\circ}\)) half of the power of the incident beam will be reflected by the wire grid at point A (see Figure 2, left). The remaining portion of the beam will travel through the grid and be reflected by the flat mirror surface at point B (see Figure 2, left).

Depending upon the distance t between the wire grid and the flat mirror surface, the portion of the beam transmitted by the wire grid will have to travel an additional distance ABC (see Figure 2, left) and will accumulate a phase difference \(\rho\) with respect to the portion of the beam taht is reflected by the wire grid at point A (see Figure 2, left).

For the incoming beam, having a wavelength of \(\lambda \), propagating at an incident angle \(\theta = 45^{\circ}\) and a grid–mirror separation of t, the optical path difference P to travel the additional distance is:

$$ P = AB + BC \\ P = 2t \cos(\theta^{\prime}). $$

The phase difference \(\rho\) between orthogonal components is given by

$$ \rho = 4 \pi t \cos(\theta^{\prime} \lambda), $$

and the separation distance between orthogonal components \(\Delta\) (beam walk-off, Figure 2 right) is given by

$$ \Delta = 2t \sin(\theta^{\prime}). $$

By changing the distance t, it is possible to change the phase difference between the two beam components and therefore to change the polarization of the e-field of the reflected beam, e.g., from linear to circular or arbitrary linear polarization orthogonal to the propagation axis, with respect to the reflected beam. Choosing the correct distance t, an incident linearly polarized THz beam can therefore be transformed into a beam with a phase difference given by \(\rho\) with respect to the orthogonal polarization. This property enables transformation of, e.g., linear polarization to circular polarization, similar to a quarter-lambda waveplate.

2.1.2 - Installation

The PTR is compatible with the Bridge12 bread-board for quasi-optical instrumentation. The bread-board is based on a 2.5 in. by 2.5 in. grid. It is commonly referred to as a 1 1/2 - 3 - 1 1/2 optic. This means, make sure that there are 2 empty cube positions between two refocusing mirrors.

Installation

To install the PTR:

  1. Slide the dowel pin into the clearance hole of the bread board
  2. Orient the PTR
  3. Lock the PTR position by fastening the 1/2-40 socket head screw

Optimizing the Position

The PTR, as all Bridge12 QO elements only have one degree of freedom (rotation about the dowl pin) to optimize the insertion loss or overall transmission loss of a QO system. To minimize transmission losses:

  1. Loosen the 1/4-20 socket head screw. Do not remove the screw from the bread board
  2. Monitor the transmitted power
  3. Rotate the PTR while maximizing the transmitted power

Note

QO systems can have many individual configurations. In some systems it is easier to monitor the transmitted power, in others it is more convenient to monitor the reflected power. If you are not sure how to optimize the transmission losses in your QO system please contact Bridge12 Technologies, Inc. at support@bridge12.com.

Warning

Do not touch the free-standing wire grid of the PTR. The grid is located at the front of the PTR. The grid is extremely fragile and can be easily destroyed by sharp objects (e.g. screw drivers).

Adapters

The PTR is fully compatible with the Bridge12 QO bread board design. Bridge12 offers adapter plates if you are using a different bread board design. Please contact us at info@bridge12.com for more information.

2.1.3 - Maintenance

The Bridge12 PTR is completely maintenance free. If you experience any problems please contact Bridge12 at support@bridge12.com.

Warning

The free-standing wire grid of the PTR is extremly fragile. Do not use any sharp tools (e.g. blades, screw drivers, etc.) when working around the PTR.

2.1.4 - References

The PTR is not a new device and has been described throughout the scientific literature. Some major references are:

  • Howard, J., W. A. Peebles, and N. C. Luhmann. “The Use of Polarization Transforming Reflectors for Far-Infrared and Millimeter Waves.” International Journal of Infrared and Millimeter Waves 7 (October 1, 1986): 1591–1603. https://doi.org/10.1007/bf01010760.
  • Earle, Keith A., Dmitriy S. Tipikin, and Jack H. Freed. “Far-Infrared Electron-Paramagnetic-Resonance Spectrometer Utilizing a Quasioptical Reflection Bridge.” Review of Scientific Instruments 67 (1996): 2502–13.
  • Amer, N., W. C. Hurlbut, B. J. Norton, Yun-Shik Lee, and T. B. Norris. “Generation of Terahertz Pulses with Arbitrary Elliptical Polarization.” Applied Physics Letters 87, no. 22 (November 28, 2005): 221111. https://doi.org/10.1063/1.2138351.
  • Chuss, David T., Edward J. Wollack, Ross Henry, Howard Hui, Aaron J. Juarez, Megan Krejny, S. Harvey Moseley, and Giles Novak. “Properties of a Variable-Delay Polarization Modulator.” Applied Optics 51, no. 2 (January 10, 2012): 197. https://doi.org/10.1364/AO.51.000197.

However, when characterizing the PTR developed by Bridge12 Technologies, Inc. we realized that the PTR can also be used as an Universal Polarizer. This properties hasn’t been reported before and was first described in:

  • Chen, Jeson, and Thorsten Maly. “Compact, Tunable Polarization Transforming Reflector for Quasi-Optical Devices Used in Terahertz Science.” Review of Scientific Instruments 93, no. 1 (January 1, 2022): 013102. https://doi.org/10.1063/5.0036292.

If you are using a PTR manufactured by Bridge12, please consider referencing the above publication.

Initial results were presented at the 2018 ENC Conference as a poster contribution by Jeson Chen.

Poster by Jeson Chen and Thorsten Maly, presented at the 2018 ENC

Poster by Jeson Chen and Thorsten Maly, presented at the 2018 ENC

3 - Accessories

Accessories for EPR/DNP Spectroscopy

3.1 - Active Electric Shims

Online documentation for the Bridge12 active electric shims for elelctromagnets

Welcome to the online documentation of the Bridge12 Active Electric Shims (AES) for electromagnets.

Bridge12 Active Electric Shims (AES)

Bridge12 Active Electric Shims (AES)

The Bridge12 AES can be installed in an electromagnet with horizontal field orientation, commonly used in EPR spectroscopy. In total 5 different coils are included to actively control the Z1, Z2, X, and Y gradients. In addition a B0 (Z0) coil is integrated to perform small field adjustments.

Please follow the installation and operation instructions. If you have further questions please contact Bridge12 at support@bridge12.com.

3.1.1 - Installation

Before You Start …

Important

You heard it many times. Before you start assembling the shims we suggest first reading the complete installation instructions to the end and make yourself familiar with the different components. To assemble the shims no further tools are required.

Warning

To install the shims, make sure the magnet is switched off.

What is Included

The Bridge12 AES come by default with the following parts (see figure below):

Bridge12 Active Electric Shims (AES) components

Bridge12 Active Electric Shims (AES) components

  1. Electrical Shims (right side, qty.: 1)
  2. Electrical Shims (left side, qty.: 1)
  3. Shim spacers (qty.: 4)
  4. Shim supports (qty.: 8)
  5. Adjustment/tightening tool (qty.: 1)
  6. Cables to connect shims to shim interface (qty.: 2)
  7. Shim interface (qty.: 1)
  8. Shim board adapter for Bruker BSMS shim power supply (qty.: 1)
  9. Cable to connect shim interface (7) to Bruker BSMS shim board adapter

Assembling the Shims

Important

Assembling the shims is best done by two people.

The electrical shims are supported between the magnet poles using a total of 8 shim supports and 4 shim spacers (shown in figure below).

Shim Supports (left, right) and Spacer

Shim Supports (left, right) and Spacer

To assemble the shims:

  1. Place the electrical shims (1 and 2) at the side of the magnet. The board labeled “left” should face the left magnet pole. The board labeled “right” should face the right magnet pole. The labels are located on the back of the boards.
  2. Place the shim spacer between the shim boards
  3. Screw in the shim supports on both side.

The assembly should look similar to the assembly shown in the figure below.

Shims with one set of supports and spacer assembled

Shims with one set of supports and spacer assembled

  1. Repeat the previous step for all shim supports and spacers.

Alternatively, you can first assemble the shim boards using the supports and the spacers outside the magnet and place the assembled shims between the magnet poles.

Under some circumstances it is not possible to use all 8 shim supports. In this case two supports can be omitted and replaced by two screws as shown in the following image. However, make sure when locking the shims in place that their position is fully secured and that the shims won’t slide down. We don’t recommend removing more than one set of shim supports.

Shims with spacer only

Shims with spacer only

Locking the Shims in Place

Once you have assembled the shims and placed between the magnet pole pieces, you need to lock the shims in place. Each support of the shims has an adjustment screw and a counter nut (see figure below). Before you start turn the counter nut to face the screw as shown in the figure below.

Shim Supports with adjustment screw and counter nut

Shim Supports with adjustment screw and counter nut

To hold the shims between the magnet poles, turn the adjustment screws on all sides, until they touch the side of the magnet. Make sure the shims are centered between the pole pieces of the magnet. To turn the screws, especially when the tops of the screws are close to the sides of the magnet use the Adjustment/Tightening tool.

Once the shims are supported between the magnet poles, lock the adjustment screws by turning the counter nut until it touches the shim support. Make sure the nut is not overtightened.

The picture below shows the shims installed in an electromagnet for EPR spectroscopy.

Picture of the installed shims (courtesy of Prof. Song-I Han, UCSB

Picture of the installed shims (courtesy of Prof. Song-I Han, UCSB

Connecting the Shims

To use the shims, the following connections have to be made:

  1. Shims to shim interface
  2. Shim interface to shim power supply
Connecting the shims to the shim interface

Connecting the shims to the shim interface

To connect the shims to the shim interface follow these steps:

  1. Starting with the shims located to the right, connect the ribbon cable marked “RIGHT” to the right shim board.
  2. Connect the same ribbon cable the shim interface.
  3. Repeat the step for the ribbon cable marked “LEFT”.

Connecting the Shims to a Lab Power Supply

The shims can be operated using any benchtop lab power supply. For convenience a multi-channel power supply can be used. Often these power supplies are not bi-polar, in which case you need to swap the leads of the respective channel if you would like to reverse the gradient (e.g. Z1 to -Z1). The power supply can be directly connected to the 15 pin sub-D connector of the shim interface.

Pin# (sub-D) Shim Connection Pin# (sub-D) Shim Connection
1 Z(B)0 (s) 2 Z(B)0 (r)
3 Z1 (s) 4 Z2 (r)
5 Z2 (s) 6 Z2 (r)
7 X (s) 8 X (r)
9 Y (s) 10 Y (r)
15 GND Shield GND
(GND - Ground, s - supply, r - return)

Connecting the Shims to a Shim Power Supply

The shims can be operated using a shim power supply e.g. of a NMR spectrometer. The following instructions are specific to a shim power supply used in Bruker NMR consoles (courtesy to Prof. Song-I Han, UCSB).

To connect the shims to a Bruker shim power supply (e.g. BSMS/2) follow this steps:

  1. Connect the shim board adapter to the the shim power supply (see image below, left).
  2. Connect the 15 pin sub-D cable to the shim interface.
  3. Connect the 15 pin sub-D cable to the shim board adapter.
Connecting the shims to the Bruker BSMS/2 shim power supply (images courtesy Prof. Song-I Han, UCSB

Connecting the shims to the Bruker BSMS/2 shim power supply (images courtesy Prof. Song-I Han, UCSB

Once you have made these connection, the shim can be controlled through the shim power supply.

If you need help at any step of the installation please contact Bridge12 at support@bridge12.com.

3.1.2 - Specifications

The Bridge12 Active Electric Shims (AES) are a 5 channel shim system for electromagnet with a horizontal field. These magnets are typically used in Electron Paramagnetic Resonance (EPR) spectroscopy. The five channels or the shims are: Z0 (or B0), Z1, Z2, X, and Y. Each channel can operate at a current of up to 1.5 A.

Parameter Value
Number of channels 5
Maximum current/channel 1.5 A
Spacing 53 mm

Warning

Do not exceed a current of (+/-) 1.5 A per channel to avoid permanent damage to the shims.

Shim Interface Pinout

Pin# (sub-D) Shim Connection Pin# (sub-D) Shim Connection
1 Z(B)0 (s) 2 Z(B)0 (r)
3 Z1 (s) 4 Z2 (r)
5 Z2 (s) 6 Z2 (r)
7 X (s) 8 X (r)
9 Y (s) 10 Y (r)
15 GND Shield GND
(GND - Ground, s - supply, r - return)

If you need help at any step of the installation please contact Bridge12 at support@bridge12.com.

3.2 - Microwave Power Source

Online documentation for the Bridge12 MPS

We are currently updating the online documentation for the Bridge12 Microwave Power Source (MPS) for ODNP spectroscopy. In the meanwhile, please visit mps.bridge12.com for the previous version.