Commonly Used OVJ Commands

General Command Structure

  • OVJ commands are entered at the command line. Some commands require arguments or parameters, which can enclosed in parentheses.
  • To change a parameter value, the parameter is entered followed by an equal sign (=), followed by the value of the parameter (real, integer, or string).
  • String values are always enclosed in parentheses and surrounded by single quotes.
  • Multiple command and parameter settings can be entered consecutively, separated by spaces.

Command Examples:

1
2
3
aph
wft
wft aph dc vsadj

Example of Setting a Parameter Value:

1
2
lb = 0.2
sampe = 'MySample'

Commonly Used OVJ Parameters

The following sections provide a list of commonly OVJ commands and parameters. For a complete list of commands see the VnmrJ Command and Parameter Reference.

Commands

Spectrometer Control and Data Acquisition Parameters
Command Description/Example
aa abort current running experiment immediately. Recommended to use in most occasions over sa except when some experiments are queued
cexp(n) create new experiment from current one
Example: cexp(2) copy current parameters and create experiment number 2 (#2 must not exist)
dg display group of acquisition/processing parameters in Process→Text Output window
dps display pulse sequence
ga start experiment and autoprocess the data when data are available
go start experiment using acquisition parameters
jexp(n) join (or go to) specified experiment
Example: jexp(2) join exp#2. exp#2 must exist
movesw move new spectral window (sw) to the area enclosed by the two red cursors
movetof move transmitter offset freqeuency (tof) to the cursor position
su setup experiment. Force hardware to changes according to current parameter settings
Example: load=‘MySpectrum’ su - change shims to current values in parameter file
Example: tn=‘C13’ su - change direct detection to C13
time displays total experiment time with current parameters
time(hours,minutes) displays number of scans (nt) required for a given time of experiment
Example: time(1,10) - displays number of scans needed for a one hour 10 mins experiment
unlock(exp_number) remove interactive lock and join an experiment. Useful to remove lockup of an experiment, sometimes due to user failure to exit vnmrJ before logging out Linux
Data Processing and Display Commands
Command Description/Example
aph automatic phasing of both zero (constant) and first-order (linear with frequency) terms
aph0 automatic phasing with zero-order term only to achieve absorptive mode spectrum. Zero-order phase value (rp) is the same across the spectrum
array Enter this command to set an arrayed experiment where a parameter is varied and an experiment is collected with each parameter. Enter parameter name, initial value, number of steps, and stepsize. All other parameters are kept constant
bc 1D or 2D baseline correction with a spline or 2nd to 20th order polynomial function. Needs to define baseline regions (regions in between the integral areas covering peaks)
centersw With full spectrum displayed and in single cursor mode, centers moves the cursor to the center (transmitter) of the spectrum
cz clear integral reset points
da display acquisition parameter array in Plot→Text Output window
dc Apply a simple drift correction (with a straight line function) so that the spectrum baseline is close to zero. Level (lvl) and tilt (tlt) parameters are adjusted
df display a single FID (default index is 1)
Example: df displays the 1st FID or df(10) displays #10 FID in the array
dli display list of integrals
dll displays line frequency and intensities (peaks have to picked)
dpir display integral values below spectrum
dpf adjust threshold and display peaks so that a maximum 20 highest peaks are found and the minimum height cutoff is 5 times above noise level
dres calculates peak resolution/linewidth
ds display processed spectrum without preprocessing
dssa display arrayed spectra in a stack plot
Example: wft dssa
dssh display arrayed spectra horizontally
Example: wft dssh
dssl Label displayed arrayed spectra with index number
Example: wft dssh dssl
dsn measure and display signal-to-noise ratio in region enclosed by two cursor lines
f display the full spectrum
fbc apply baseline correction for each spectrum in an spectrum array
full display spectrum with full window size. The range of the spectrum is not changed
Example: f full - display the full spectrum with full window size
fp find peaks in arrayed dataset according to the refrence peaks found by dll
isadj automatic adjustment of integral vertical display height to fit page
mf move FID file from one experiment to another
Example: mf(1,2) move fid from exp #1 to #2
mp Move all parameters from one experiment to another
Example: mp(1,2)
nl move cursor line to nearest peak/line position
th set a threshold for peaks in dll command
thadj(maxpks,noise_mult) Adjust the threshold th so that no more than maxpks number of peaks are found with a minimum of noise_mult factor above noise level.
Example: thadj(20,5)
peak Find tallest peak in current display region
Example: peak - display tallest peak position and height in current displayed spectrum region
Example: peak:$height,$freq - find tallest peak in current displayed region and save height and frequency in the variables $height and $freq. $heigt? or $freq? shows the values
vsadj automatic vertical scale adjustment so that the highest peak fits the window height
setref set reference ppm of any nucleus by the chemical shift of the deuterium signal from the solvent, according to IUPAC standard indirect referencing to TMS 1H signal
wft Weighted Fourier transform (FT). Weighting is applying a window function to the profile of the FID to enhance resolution or signal-to-noise and to reduce truncation artifacts from finite data collection

Plotting Commands

Command Description/Example
page submit plot to printer and change plotter page. Usually last command in printing command chain
pap plot “all” parameters in table format
pir plot integrals below spectrum
pl plot spectrum
pll plot frequency line list in a table format
pltext plot user comment/notes. Not necessary if ppa command is used
ppa plot basic parameters in plain english format
pps plot pulse sequence
ppf plot peak frequencies over the spectrum

Parameters

Acquisition parameters

Parameter Description/Example
at acquisition time in seconds
bs block size of data (number of transients) periodically read and stored on disk
Example: bs=4 stores data every 4 scans
ct current transient (or scan) number. ct is displayed at the bottom of vnmrJ screen during a running experiment. If the experiment is stopped before all transients are done, ct (stored in the parameter file procpar) is the number of scans completed
d1 1st delay period in pulse sequence. Typically this is the recycle time (in seconds) between scans
dn 1st decoupler nucleus
gain receiver gain. If gain=‘n’, auto-gain adjustment is set before data acquisition. Autogain cannot be used for arrayed experiment. Gain values go from 0 to 60. Too high a gain setting causes receiver overflow, leading to severe artifacts in the spectrum
Example: gain=36 sets gain to 36
nt number of transients or scans
pad preacquisition delay (in seconds). pad is the additional delay time set before the 1st recycle delay (d1) before the start of an experiment. If pad=60, the experiment will start after 60 secs
rattn receiver attenuation, the attenuation can be set from 0 to 95, a too high value will attenuate all signals. A too small value will increase noise without increasing the signal strength.
ss number of steady-state transients or dummy scans before start of experiment. It’s used to establish a steady-state for the spins before data collection
sw spectral width of direct detection dimension, in Hz be default
Example: sw=6000 or sw=15p to specify 15ppm width
sw1 spectral width of 1st indirect detection dimension, in Hz be default
solvent name of solvent
Example: solvent=‘cdcl3’ or solvent=‘c6d6’
tn observe transmitter nucleus
tof observe transmitter offset (center of spectrum)

Processing and display parameters

Parameter Description/Example
axis axis label in display and plots. Values include ‘h’ for Hz, ‘p’ for ppm
Example: axis=‘p’ or axis=‘ppm’ set axis to ppm display
cr cursor position in direct detected dimension
Example: cr=8.0p set cursor position to 8ppm
date date of experiment
delta In two-cursor (box cursor) mode, delta stores the width in Hz between the two cursors
Example: delta=1000 sets the separation between two cursors to 1000 Hz, or delta? to see the cursor separation value
ho horizontal offset (in mm) between a set of spectra in stacked display mode for arrayed spectra
intmod integral display mode. Value is ‘off’, ‘full’, or ‘partial’
Example: intmod=‘off’
lb line broadening amount for exponential weighting function (Hz)
Example: lb=0.2
lp 1st order (or linear) phase in directly detected dimension, adjusted during phasing process
Example: lp=0 sets linear phase correction to zero, useful to reset linear phase to zero to correct accidental introduction of large lp value during manual phasing
rp zero order phase correction in degree, adjusted during phasing process
Examples: rp=45 set zero-order phase to 45 degree or rp=rp+45 to change phase by 45 degree
vo vertical offset (in mm) between a set of spectra in stacked display mode for arrayed spectra
vp vertical position of spectrum, in mm
Example: vp=60 sets spectrum baseline roughly in the middle of the display along Y-axis
Example: vp=vp+20 moves the spectrum up by 20 mm
vs vertical scale of spectrum
Example: vs=vs*2 - set vertical scale to double peak height
Example: vs=vs*0.5 - set vertical scale to halve peak height